EXPHIL EXPRESS

(800) 282-5632

 
tds3012 tds3014 tds3032 tds3034 tds3052 tds3054
MSO4000 Series Mixed Signal Scopes
Specifications
  Home  |  Products  |  Price quote  |  Contact us  |  Search

 Tektronix MSO4000 Series Specifications.....


Download MSO4000 Series Data Sheet

Tektronix 4000 Series Family • DPO4000 Series • MSO4000 Series  Specifications

Vertical System 

Analog Channels 

 DPO4032 

 MSO4032 

 DPO4034 

 MSO4034 

 DPO4054 

 MSO4054 

 DPO4104 

 MSO4104 

Input Channels 2 4 4 4
Analog Bandwidth (-3 dB) 5 mV/div - 1 V/div 350 MHz 350 MHz 500 MHz 1 GHz
Calculated Rise Time 5 mV/div (typical) 1 ns 1 ns 700 ps 350 ps
Hardware Bandwidth Limits 20 MHz or 250 MHz
Input Coupling AC, DC, GND
Input Impedance 1 MΩ ±1%, 50 Ω ±1%
Input Sensitivity, 1 MΩ 1 mV/div to 10 V/div
Input Sensitivity, 50 Ω 1 mV/div to 1 V/div
Vertical Resolution 8 bits
Max Input Voltage, 1 MΩ 250 VRMS with peaks ≤ ± 400 V
Max Input Voltage, 50 Ω 5 VRMS with peaks <±20 V
DC Gain Accuracy ±1.5% with offset set to 0 V
Offset Range 1 MΩ 50 Ω
1 mV/div to 50 mV/div ±1 V ±1 V
50.5 mV/div to 99.5 mV/div ±0.5 V ±0.5 V
100 mV/div to 500 mV/div ±10 V ±10 V
505 mV/div to 995 mV/div ±5 V ±5 V
1 V/div to 5 V/div ±100 V ±5 V
5.05 V/div to 10 V/div ±50 V NA
Channel-to-channel Isolation Any Two Channels at Equal Vertical Scale ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth
                   
Vertical System 

Logic Channels 

MSO4032 MSO4034 MSO4054 MSO4104
Input channels 16 Digital (D15 - D0)
Thresholds Per Channel Thresholds
Threshold Selections TTL, CMOS, User Defined
User Defined Threshold Range +5 to -2V
Maximum Input Voltage ±15 V
Threshold Accuracy ±(100 mV + 3% of threshold setting)
Input Dynamic Range 6 Vp-p centered around the threshold
Minimum Voltage Swing 500 mV
Input Impedance 20 kΩ
Channel-to-Channel Skew 1 ns Typical
Probe Loading 3 pF
Vertical Resolution 1 bit

 

 

Horizontal System 

Analog Channels 

 DPO4032 

 MSO4032 

 DPO4034 

 MSO4034 

 DPO4054 

 MSO4054 

 DPO4104 

 MSO4104 

Maximum Sample Rate (all channels) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s
Maximum Record Length (all channels) 10 M points
Maximum Duration at Highest Sample Rate (all channels) 4 ms 4 ms 4 ms 2 ms
Timebase Range 1 ns to 1,000 s 400 ps to 1,000 s
Timebase Delay Time Range -10 divisions to 50 s
Channel-to-channel Deskew Range ±100 ns
Timebase Accuracy ±5 ppm over any ≥ 1 ms interval
                   
Horizontal System

Logic Channels 

MSO4032 MSO4034 MSO4054 MSO4104
Maximum Sample Rate (Main) 500 MS/s
Maximum Record Length (Main) 10 M points
Maximum Sample Rate (MagniVu) 16.5 GS/s
Maximum Record Length (MagniVu) 10 K points centered on the trigger
Minimum Detectable Pulse Width 1.5 ns

 

Trigger System

Main Trigger Modes - Auto, Normal, and Single.

Trigger Coupling - DC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity).

TriggerHoldoff Range - 20 ns to 8 s.

Sensitivity

Internal DC Coupled - 0.4 div DC to 50 MHz, increasing to 1 div at rated bandwidth.

External (Auxiliary Input) - 200 mV from DC to 50 MHz increasing to 500 mV at 250 MHz.

Trigger Level Range

Any Channel - ±8 divisions from center of screen.

External (auxiliary input) - ±8V.

Trigger Modes

Edge - Positive or negative slope on any channel or front panel auxiliary input. Coupling includes DC, HF reject, LF reject and noise reject.

Sequence (B-trigger) - Trigger Delay by Time - 4 ns to 8 s. Or Trigger Delay by Events - 1 to 9,999,999 events.

Pulse Width - Trigger on width of positive or negative pulses that are >, <, = or ≠ a specified period of time.

Runt - Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.

Logic - Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock to look for the pattern on a clock edge.
Pattern (AND, OR, NAND, NOR) specified for all analog and digital input channels defined as High, Low or Don’t Care.

Setup and Hold - Trigger on violations of both setup time and hold time between clock and data present on any of the input channels.

Rise/Fall Time - Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative or either.

Video - Trigger on line number, all lines, odd, even or all fields on NTSC, PAL and SECAM video signals.

I2C (Optional) - Trigger onStart, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data or Address & Data on I2C buses up to 3.4 Mb/s.

SPI (Optional, 4-channel models only) - Trigger on SS, MOSI, MISO or MOSI & MISO on SPI buses up to 50.0 Mb/s.

CAN (Optional) - Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data, End of Frame
or Missing ACK on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤, <, =, >, ≥ or ≠ a specific data value. User-adjustable sample point is set to 50% by default.

RS-232 (Optional) - Trigger on Tx start bit, Rx start bit, Tx end of packet, Rx end of packet, Tx data, Rx data.

Parallel (available on MSO models only) - trigger on a parallel bus data value.

Acquisition Modes

Sample - Acquires sampled values.

Peak Detect - Captures narrow glitches at all real-time sampling rates.

Averaging - From 2 to 512 waveforms included in average.

Envelope - Min-max envelope reflecting Peak Detect data over multiple acquisitions.

Hi-Res - Real-time boxcar averaging reduces random noise and increases vertical resolution.

Roll - Scrolls waveforms right to left across screen at sweep speeds slower than or equal to 40ms/div.

Waveform Measurements

Cursors - Waveform and Screen.

Automatic Measurements - 25, of which up to four can be displayed on screen at any one time.
Measurements include Period, Frequency, Delay, Rise Time, Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width,
Burst Width, Phase, Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Area and Cycle Area.

Measurement Statistics - Mean, Min, Max, Standard Deviation.

Reference Levels - User-definable reference levels for automatic measurements can be specified in either percent or units.

Gating - Isolate the specific occurrence within an acquisition to take measurements, using either the screen or waveform cursors.

WaveformMath

Arithmetic - Add, subtract, multiply, and divide waveforms.

Math Functions - Integrate, Differentiate, FFT.

FFT - Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or Blackman-Harris.

Advanced Math - Define extensive algebraic expressions including analog waveforms, math functions, scalars, up to two user adjustable variables and results
of parametric measurements e.g., (Intg (Ch1–Mean(Ch1)) x 1.414 x VAR1).

Software

National Instruments SignalExpress Tektronix Edition - A fully interactive measurement software environment optimized for the 4000 Series, enables you to instantly acquire, generate, analyze, compare, import and save measurement data and signals using an intuitive drag-and-drop user interface that does not require any programming. Standard 4000 Series support for acquiring, controlling, viewing and exporting your live signal data is permanently available through the software. The Professional version adds additional signal processing, advanced analysis, mixed signal, sweeping, limit testing and user-defined step capabilities and is available for a 30-day trial period standard with each instrument.

OpenChoiceź Desktop - Enables fast and easy communication between a Windows PC and the 4000 Series, via USB or LAN. Transfer and save settings, waveforms, measurements and screen images.

IVI Driver - Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI, Microsoft .NET and MATLAB.

Display Characteristics

DisplayType - 10.4 in. (264mm) liquid crystal TFT color display.

DisplayResolution - 1,024 horizontal x 768 vertical pixels (XGA).

WaveformStyles - Vectors, Dots, Variable Persistence, Infinite Persistence.

Input/Output Ports

CompactFlash Drive - Front panel access (Type 1).

USB2.0 Full Speed Host Port - Supports USB mass storage devices, printers and keyboard. Two ports available on rear-panel and one on front-panel.

USB 2.0 High Speed Device Port - Rear panel connector allows for control of oscilloscope via USBTMC or GPIB with a TEK-USB-488.

LAN Port - RJ-45 connector, supports 10/100Base-T.

XGA VideoPort - DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector.

Auxiliary Input - Front panel BNC connector. Input Impedance 1MΏ. Max input 250 VRMS with peaks ±400 V.

Probe Compensator Output - Front panel pins. Amplitude 2.5 V. Frequency 1 kHz.

Trigger Out - Rear-panel BNC connector, provides a positive polarity pulse when the oscilloscope triggers.

Kensington Lock - Rear-panel security slot connects to standard Kensington lock.

PowerSource

Power Source Voltage - 100 to 240 V ±10%.

PowerSource Frequency - 47 to 66 Hz (90 to 264 V), 360 to 440 Hz (100 to 132 V).

Power Consumption - 250W maximum.

Physical Characteristics

 

Dimensions mm in
Height 229 9.0
Width 439 17.3
Depth 137 5.4
Weight kg lbs
Net 5 11
Shipping 9.5 22
Rackmount Configuration 5U

 

General Characteristics

Cooling Clearance - 2 in.(51mm) required on left side and rear of instrument.

Environmental

Temperature

Operating - 0șC to +50șC.

Non-operating -  –20șC to +60șC.

Humidity

Operating -

High: 40șC to 50șC, 10% to 60% Relative Humidity

Low: 0șC to 40șC, 10% to 90% Relative Humidity

 

Non-operating -

High: 40șC to 60șC, 5% to 60% Relative Humidity

Low: 0șC to 40șC, 5% to 90% Relative Humidity

 

Altitude

Operating - 3,000 meters (9,843 feet).

Non-operating - 12,000 meters (39,370 feet).

Random Vibration

Operating - 0.31 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total.

Non-operating - 2.46 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total.

Regulatory

Electromagnetic Compatibility - 89/336/EEC Safety – UL61010-1, Second Edition; CSA61010-1 Second Edition, EN61010-1:2001; IEC 61010-1:2001.

Download MSO4000 Series Data Sheet


End of the Tour NEXTARROWGIF.GIF (159 bytes)    Return to MSO4000 Main Page

  Home  |  Products  |  Price quote  |  Contact us   |  Search

1-800-282-5632