EXPHIL EXPRESS

(800) 282-5632

 
tds3012 tds3014 tds3032 tds3034 tds3052 tds3054
DPO3000 Series Specifications
  Home  |  Products  |  Price quote  |  Contact us  |  Search

 Tektronix DPO 3000 Series Specifications.....
Vertical System           DPO3012 DPO3014 DPO3032 DPO3034 DPO3052 DPO3054
Input Channels 2 4 2 4 2 4
Analog Bandwidth (-3dB) 100 MHz 100 MHz 300 MHz 300 MHz 500 MHz 500 MHz
Calculated Rise Time

5 mV/div (typical)

3.5 ns 3.5 ns 1.17 ns 1.17 ns 700 ps 700 ps
Hardware Bandwidth Limits 20 MHz or 150 MHz
Input Coupling AC, DC, GND
Input Impedance 1 MΩ ±1%, 75 Ω ±1%, 50 Ω ±1%
Input Sensitivity Range, 1 MΩ 1 mV/div to 10 V/div
Input Sensitivity Range, 75 Ω, 50 Ω 1 mV/div to 1 V/div
Vertical Resolution 8 bits (11 bits with Hi-Res)
Max Input Voltage, 1 MΩ 300 VRMS with peaks ≤±450 V
Max Input Voltage, 75 Ω, 50 Ω 5 VRMS with peaks ≤± 20 V
DC Gain Accuracy ±1.5% with offset set to 0 V
Offset Range                       1 MΩ 50 Ω, 75 Ω
   1 mV/div to 99.5 mV/div ±1 V ±1 V
   100 mV/div to 995 mV/div ±10 V ±5 V
   1 V/div ±100 V ±5 V
   1.01 V/div to 10 V/div ±100 V NA
Channel-to-Channel Isolation

(Any Two Channels at Equal Vertical Scale)

≥ 100:1 at ≤100 MHz and ≥ 30:1 at > 100 MHz up to the rated BW

 

Horizontal System All DPO3000 Models
Maximum Sample Rate (all channels) 2.5 GS/s
Maximum Record Length (all channels) 5 M points
Maximum Duration of Time Captured at Highest Sample Rate (all channels) 2 ms
Time base Range (S/div) 1 ns to 1000 s
Time base Delay Time Range - 10 divisions to 5000 s
Channel-to-Channel Deskew Range ±100 ns
Timebase Accuracy ±10 ppm over any ≥ 1 ms interval

 

Trigger System
Main Trigger Modes Auto, Normal, and Single
Trigger Coupling DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)
Trigger Holdoff Range 20 ns to 8 s
Sensitivity
Internal DC Coupled 0.4 div DC to 50 MHz, increasing to 1 div at rated bandwidth
External (Auxiliary Input) 200 mV from DC to 50 MHz increasing to 500 mV at 250 MHz
Trigger Level Range
Any Channel ±8 divisions from center of screen
External (auxiliary input) ±8 V
Acquisition Modes
Sample Acquire sampled values
Peak Detect Captures narrow glitches at all real-time sampling rates
Averaging From 2 to 512 waveforms included in average
Envelope Min-max envelope reflecting Peak Detect data over multiple acquisitions
Hi-Res Real-time boxcar averaging reduces random noise and increases resolution
Roll Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div
Trigger Modes
Edge Positive or negative slope on any channel or front panel auxiliary input. Coupling includes DC, AC, HF reject, LF reject and noise reject
Pulse Width Trigger on width of positive or negative pulse that are >, <, = or ≠ a specified period of time
Runt Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again
Logic Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for four input channels defined as High, Low or Don’t Care.
Setup and Hold Trigger on violations of both setup time and hold time between clock and data present on any two input channels
Rise/Fall Time Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative or either.
Video Trigger on all lines, odd, even or all fields on NTSC, PAL and SECAM video signals
Extended Video (optional) Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/30, 1080p/50, 1080p/60 and custom bi-level and tri-level sync video standards
I2C (optional) Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data or Address & Data on I2C buses up to 3.4 Mb/s
SPI (optional) Trigger on SS, MOSI, MISO or MOSI & MISO on SPI buses up to 10.0 Mb/s
CAN (optional) Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data, End of Frame, Missing ACK, or Bit Stuffing Error on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤, <, =, >, ≥ or ≠ a specific data value. User-adjustable sample point is set to 50% by default.
LIN (optional) Trigger on Sync, Identifier, Data, Id & Data, Wakeup Frame, Sleep Frame, Error up to 100 kb/s
RS-232/422/485/UART (optional) Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data, Tx Parity Error, and Rx Parity Error
Trigger Delay by Time 4 ns to 8 s
Trigger Delay by Events 1 to 9,999,999 events

 

Waveform Measurements
Cursors Waveform and Screen
Automatic Measurements 29, of which up to four can be displayed on screen at any one time.

Measurements include: Period, Frequency, Delay, Rise Time, Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Rising Edge Count, Falling Edge Count, Positive Pulse Count, Negative Pulse Count, Area and Cycle Area.

Measurement Statistics Mean, Min, Max, Standard Deviation
Reference Levels User-definable reference levels for automatic measurements can be specified in either percent or units
Gating Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, or waveform cursors
Waveform Math
Arithmetic Add, subtract, multiply, and divide waveforms
Math Functions Integrate, Differentiate, FFT
FFT Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or Blackman-Harris
Advanced Math Define extensive algebraic expressions including waveforms, math functions, scalars, up to two user adjustable variables and results of parametric measurements e.g.,(Intg(Ch1–Mean(Ch1)) x 1.414 x VAR1)

Software

NI LabVIEW SignalExpressTM Tektronix Edition A fully interactive measurement software environment optimized for the DPO3000 Series, enables you to instantly acquire, generate, analyze, compare, import and save measurement data and signals using an intuitive drag-and-drop user interface that does not require any programming.

Standard DPO3000 Series support for acquiring, controlling, viewing and exporting your live signal data. A 30-day trial period of the Full Version provides additional signal processing, advanced analysis, mixed signal, sweeping, limit testing and user-defined step capabilities. Order SIGEXPTE for permanent Full Version capability.

OpenChoice® Desktop Enables fast and easy communication between a Windows PC and the DPO3000 Series, via USB or LAN. Transfer and save settings, waveforms, measurements and screen images.
IVI Driver Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI, Microsoft .NET and MATLAB

Display Characteristics

Display Type 9 in. (228.6 mm) wide format liquid crystal TFT color display
Display Resolution 800 horizontal x 480 vertical pixels (WVGA)
Waveform Styles Vectors, Dots, Variable Persistence, Infinite Persistence
Graticules Full, Grid, Cross Hair, Frame, IRE and mV
Format YT and XY
Waveform Capture Rate Up to 50,000 wfm/s

Input/Output Ports

USB 2.0 High Speed Host Port Supports USB mass storage devices and printers. One port available on rear-panel and one on front-panel.
USB 2.0 High Speed Device Port Rear panel connector allows for control of oscilloscope via USBTMC or GPIB with a TEK-USB-488
LAN Port RJ-45 connector, supports10/100Base-T
Video Out Port DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector
Auxiliary Input Front panel BNC connector. Input Impedance 1MΩ. Max input 300VRMS Cat II with peaks ≤± 450 V
Probe Compensator Output Front panel pins. Amplitude 2.5 V. Frequency 1 kHz.
Trigger Out Rear-panel BNC connector, provides a positive polarity pulse when the oscilloscope triggers
Kensington Lock Rear-panel security slot connects to standard Kensington lock

Power Source

Power Source Voltage 85 to 265 V ±10%
Power Source Frequency 45 to 440 Hz (85 to 265 V)
Power Consumption 120 W maximum
Optional TekVPI® Power Supply Output Voltage – 12 V; Output Current – 5A; Power Consumption – 60 W

Physical Characteristics

Dimensions mm in.
Height 203.2 8
Width 416.6 16.4
Depth 137.2 5.4
Weight kg lbs.
Net 4.17 9.2
Shipping 8.62 19
Rackmount Configuration 5U
Cooling Clearance 2 in. (51mm) required on left side and rear of instrument

  General Characteristics

Environmental    
Temperature    
   Operating 0ºC to +50ºC
   Non-operating –40ºC to +71ºC
Humidity    
   Operating High: 30ºC to 50ºC,5% to 45% Relative Humidity

Low: 0ºC to 30ºC,5% to 95% Relative Humidity

   Non-operating High: 30ºC to 50ºC,5% to 45% Relative Humidity

Low: 0ºC to 30ºC,5% to 95% Relative Humidity

Altitude    
   Operating 3,000 meters (9,843 feet)
   Non-operating 12,000 meters (39,370 feet)
Random Vibration    
   Operating 0.31 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total
   Non-operating 2.46 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total
Regulatory    
Electromagnetic Compatibility 89/336/EEC
Safety UL61010-1,Second Edition; CAN/CSA C22.2 No. 1010.1 1992, EN61010-1:2001; IEC 61010-1:2001