Measured Components |
L(Inductance),C(Capacitance), R(Resistance) |
|
Measurement Circuit configurations |
Series and Parallel |
|
|
|
|
|
Measured Parameter |
R+Q |
|
Measurement Ranges |
R: 10µW
to 100MW
Q: 0.0001 to 9,999 |
|
|
|
|
|
Measured Parameter |
L+Q |
|
Measurement Range |
L: 0.01µH
to 99,999 H
Q: 0.0001 to 9,999 |
|
|
|
|
|
Measured Parameters |
C+D and C+R |
|
Measurement Range |
C:
0.00001pF to 99,999µF
D: 0.0001 to 9,999
R: 0.0001W to 9,999KW |
|
|
|
|
|
Display Type |
320x 240 Graphics LCD W/Backlight |
|
Display Value |
5 digit LCR
and a 4 digit Q/D measurement |
|
Parameters Displayed |
Measured values, absolute D
and %D |
|
|
|
|
|
Test Frequencies |
500 test frequencies from 12Hz to 100KHz |
|
Test Voltage |
5mV to
1.275V in 5mV increments |
|
Test Voltage Accuracy |
±(5% + 5mV) (1 + 001 x F2) |
|
Measurement rates |
Slow: 1
Meas/sec
Medium: 5 Meas/sec
Fast: 12 Meas/sec |
|
Measurement Delay |
from 1ms to 99,999ms |
|
Ranging |
Auto or
Manual. |
|
DC Bias |
Internal: 2V, External: 0 to 60V |
|
Triggering |
Continuous
or Triggered |
|
Zeroing Correction |
Open and short circuit |
|
Averaging |
1 to 255
Avgs |
|
|
|
|
|
Basic Accuracy |
± >0.02% at 23°C ±5°C (73°F ±9°F)
(See user manual for all accuracy conditions) |
|
Remote Operation |
GPIB and
Handler (Optional) |
|
|
|
|
|
AC Voltage input |
120 or 220 Volts |
|
Frequency |
50/60 Hz |
|
Power Consumption |
20Watts (Max) |
|
Operating Temperature |
0 to 50°C
(32 to 122°F) @ relative humidity < 85% |
|
Size |
5.5" (H) x 15" (W) x 18" (D) |
|
Weight |
17.6 lbs |
|
Accessories |
Manual, Line Cord, Kelvin leads with bag, GPIB
Handler (Optional) |
|