| VERTICAL SYSTEM |
| Sensitivity
and Accuracy |
5mV/DIV to
5V/DIV +3%, 1mV/DIV to 1V/DIV +5%(using x 5Mag.) |
| Bandwidth(-3dB) |
DC to
20MHz(x 5mag.:DC to 7MHz) |
| Rise Time |
Approx.
17.5nS (x 5Mag.:Approx. 50nS) |
| Max. Input |
300V (DC+AC
peak) at 1kHz |
| Input
Coupling |
AC, GND, DC |
| Input
Impedance |
Approx.
1Mohm/25pF |
| Vertical
Mode |
CH1, CH2,
DUAL (ALT, CHOP), ADD |
| Polarity
Invert |
CH2 only |
|
|
| HORIZONTAL SYSTEM |
| Sweep Time |
0.2µS/DIV
to 0.5S/DIV |
| Accuracy |
+3%,
x 10Mag :+5% |
| Max. Sweep
Rate |
100nS/DIV
(20nS and 50nS/DIV are uncalibrated) |
|
|
| TRIGGER SYSTEM |
| Trigger
Mode |
AUTO, NORM,
TV-V, TV-H |
| Trigger
Source |
CH1,
CH2, LINE, EXT, TRIG-ALT |
| Trigger Sensitivity |
| Frequency |
INT |
ALT |
EXT |
| 20Hz ~
2MHz |
0.5 DIV |
2 DIV |
200 DIV |
| 2Hz ~
20MHz |
1.5 DIV |
3 DIV |
800 DIV |
|
| TV Trigger
Sensitivity |
more than 1
DIV(EXT:1V) |
| Trigger
Slope |
"+"
or "-" |
| EXT Trigger
Input |
Impedance:Approx.1Mohm/30pF,
Max. input voltage: 300V (DC+AC peak) at 1kHz |
|
|
| X-Y OPERATION |
| X-axis Input |
CH1 |
| Y-axis
Input |
CH2 |
| Bandwidth |
x-axis:DC
to 500kHz (-3dB) |
| Phase
Difference |
3° or less
from DC to 50kHz |
|
|
| Z-AXIS INPUT |
| Sensitivity |
At least 5Vp-p |
| Polarity |
Positive-going
input decreases |
| Frequency range |
DC to 2MHz |
| Max. Input
Voltage |
30V(DC + AC
peak) |
|
|
| CRT |
| Type |
Large
6-inch rectangular type with internal graticule |
| Effective
Area |
8 x 10 DIV
(1div=10mm) |
| Acceleration
potential |
Approx. 2kV |
| Trace
Rotation |
Provided |
| Focus
Control |
Auto focus |
|
|
| OTHERS |
| Environment
Temperature |
0 ~
40°C(32 ~ 104°F), 85% RH, Rated range of use: 10 ~ 35°C(50 ~ 95°F), 85% RH |
| Calibrator |
Frequency:
Approx. 1kHz square wave, Voltage: 2V+2% |
| CH1 Output |
Voltage:
20mV/DIV or more,Bandwidth: 50Hz ~ 5MHz (-3dB), Impedance: Approx.50ohm |
| Power
Supply |
115/230V AC+15%,
50/60Hz |
| Power
Consumption |
Approx.
40W, 45VA |
| Dimensions |
310(W) x
150(H) x 455(D)mm |
| Weight |
Approx.
8kgs(17.6 lbs) |
| Accessories |
Detachable
power cord x 1; Low capacitace probe (1:1, 1:10) x 2, Instruction manual x1 ; GTL-101 x 1 |