| Vertical Deflection | Bandwidth:
Rise time:
Bandwidth limiter: User selectable 10 kHz or 20 MHz Number of inputs: 2 plus external trigger. All isolated from each other and
ground. Input coupling: AC, DC with ground level indicator Input sensitivity: 5 mV/div to100 V/div Input voltage: See general specifications for maximum rating Vertical resolution: 8 bit Accuracy: ±(1.5% of reading + 0.04 x range/div) Input impedance: 1 MW ± 1% // 15 pF ± 2 pF |
| Horizontal | Maximum real-time sample
rate:
Number of digitizers: 2 Time base range:
Maximum record length: 1,000 points per input Accuracy: ± (0.01% of reading + 1 pixel) Glitch capture: 50 nsec (5 µsec/div to 1 min/div) Faster time bases have higher
sample rates than |
| Display and acquisition | Display modes: Input A,
input B, dual, average, persistence, invert, replay Acquisition modes: Normal, auto, single shot, ScopeRecord, roll, glitch
capture |
| Trigger and delay | Source: Input A, input B,
external trigger input. All input references isolated from each other and ground. Modes: Automatic Connect-and-View, free run, single shot, edge, delay,
video, video line, selectable pulse width Connect-and-View: Advanced automatic triggering that recognizes signal
patterns, automatically sets up and continuously adjusts triggering, time base and
amplitude. Automatically displays stable waveforms of complex and dynamic signals like
motor drive and control signals. Video triggering: NTSC, PAL, PAL+, SECAM. Includes field 1 and 2 and line
select. Pulse width triggering: Pulse width qualified by time. Allows for triggering Time delay: 9 divisions pre-trigger view to 1,000 divisions trigger delay |
| Automatic capture of 100 screens | The instrument ALWAYS memorizes
last 100 screens (no user interaction or setup required). When an anomaly occurs on
screen, theres 10 seconds to press HOLD and review it. If one sets up the instrument
for triggering on glitches or intermittent anomalies the unit operates in baby-sit
mode and will capture 100 events. Replay: Manual or continuous replay. Displays the captured 100 screens as a
live animation. The contents can also be viewed by manually scrolling
backwards and forwards screen by screen. Replay storage: Up to 2 sets of 100 screens can be saved for later recall and
analysis. |
| Automatic scope measurements | VDC, VAC rms, VAC+DC, Vpeak max,
Vpeak min, Vpeak to peak, frequency (Hz), positive pulse width, negative pulse width,
positive duty cycle, negative duty cycle, amp AC, amp DC, amp AC+DC, power factor, watts,
VA, VA reactive, phase, temperature °C, temperature °F, dBV, dBm into 50W and 600W |
| Cursor Measurements | Source: Input A or B Dual horizontal lines: Voltage at cursor 1 and 2, voltage between cursors Dual vertical lines: Time between cursors, voltage between markers Single vertical line: Min-Max and Average voltage at cursor position |
| Zoom | Up to 8x horizontal zoom |
| Meter Mode | |
| General | Via 4 mm banana inputs. Fully
isolated from scope inputs and scope ground. The specified accuracy is valid over the
temperature range 18 °C to 28 °C (15 °F to 33 °F). Add 10 % of specified accuracy for
each degree °C below 18 °C or above 28 °C (15 °F to 33 °F). |
| Maximum resolution | 5,000 counts |
| Voltmeter ranges | 500mV, 5V, 50V, 500V, 1,000V |
| Accuracy | VDC: ± (0.5 % + 5
counts) VAC true rms VAC+DC true rms |
| Ohms | Ranges: 500W, 5kW, 50kW,
500kW, 5MW, 30MW Accuracy: ± (0.6 % + 5 counts) |
| Other meter functions | Continuity: Beeper on
< 30W (± 10W) Diode test: Up to 2.8V Amps: Amp DC, Amp AC, Amp AC+DC using an Temperature (°C, °F): With optional accessories. Input impedance: 1 MW ± 1% // 10 pF ± 2 pF Advanced meter functions: Auto/manual ranging, relative measurements (Zero
reference), TrendPlot recording |
| Recorder Mode | |
| ScopeRecord-roll mode | Dual input waveform storage
mode. Source and display: Input A, Input B, Dual Memory depth: 27,500 points per input. Each point consist of Min-Max pair. Min-Max values: Min-Max values are measured at high sample rate ensuring capture and display of glitches Sample Recording Settings:
Set Time base range = 2 min/div:
Recording modes: Single sweep, continuous roll, start on external trigger Horizontal scale: Time from start, time of day Zoom: Up to 100 x zoom Memory: Up to 2 dual input ScopeRecordings can be saved for later recall and
analysis |
| TrendPlot Recording | Dual input electronic paperless
chart recorder. Plots, displays and stores meter and scope measurements. Source and display: Input A, Input B and DMM input Memory depth: 13,500 points record per input. Per record point a minimum,
maximum and average value, and a time and date stamp are stored. Ranges: 10 s/div to 20 min/div in normal view mode. 10 min/div to 24 hour/div in
view-all mode (overview of total record) Recorded timespan: Up to 8 days with a resolution of 1 minute Recording mode: Continuous roll Measurement speed: 2.5 measurements per second maximum Horizontal scale: Time from start, time of day Zoom: Up to 64x zoom Memory: Up to 2 TrendPlot recordings can be saved for later recall and analysis |
| Cursor measurements - All Recorder Modes | Source: Input A, B or DMM
input Dual vertical lines: Min-Max or Average voltage. Time between cursors. Single vertical line: Min-Max or Average voltage. Absolute date and time or time
from start. |
| General Specifications | |
| Case | Design: Rugged, shock
proof with integrated protective holster Drip and dust proof: IP51 according to IEC529 Shock and vibration: Shock 30g, Vibration 3g |
| Display | Bright LCD with CCFL backlight,
35/60 cd/m2 without/with power adapter. Size: 105 x 79 mm (4.1 x 3.1 inches) Resolution: 240 x 240 pixels Contrast and brightness: User adjustable, |
| Memory Save and Recall | Scope memories: 10 memory
locations that each can contain two waveforms plus corresponding setup. Recorder memories: 2 memory locations that each can contain 100 captured dual input scope screens, or a dual input ScopeRecord (27,500 Min-Max pairs per input), or a dual input Trendplot (13,500 Min-Max pairs per input). |
| Real-time Clock | Time and date stamp for ScopeRecord, 100 captured screens and Trendplots. |
| Power | Line power: Country
specific line voltage adapter/battery charger included Battery power: Rechargeable NiMH (installed) Battery operating time: 4 hours Battery charging time: 4 hours Battery power saving functions: Auto power down with adjustable power down time.
On screen battery power indicator |
| Mechanical data | Size: 256 x 169 x 64 mm
(10.1 x 6.6 x 2.5 inches) Weight: 1.95 kg (4.3 lbs) |
| Safety | Compliance: EN61010-1
(1993) Pollution degree 2, UL 3111-1 (1994), CAN/CSA C22.2 No.1010.1 (1992), ANSI/ISA S82.01 (1994) |
| Input voltage ratings | Maximum probe voltage:
1,000V CAT II, 600V CAT III (Maximum voltage between 10:1 probe tip (VP190) and reference lead) Floating voltage: 1,000V CAT II, 600V CAT III Independently isolated: 1,000V CAT II, 600V CAT III Maximum voltage on BNC input directly Maximum voltage on meter input: 1,000V CAT II, 600V CAT III |
| Environmental | Operating temperature: 0
°C to +50 °C Storage temperature: -20 °C to +60 °C Humidity Maximum operating altitude: 3,000 m (10,000 feet) Maximum storage altitude: 12 km (40,000 feet) Electro-Magnetic Compatibility: EN 61326-1 for emission and immunity |
| Optically isolated PC/Printer interface | To printer: Supports HP
Laserjet®,Deskjet®,Epson FX/LQ® and postscript
printers via optional PAC 91 To PC: Transfer instrument settings, screen images and waveform data, compatible with FlukeView®software for Windows® via optional PM9080. |
| Warranty | 3 years |