Measurement parameters |
|
3532-50 LCR HiTester |
3522-50 LCR HiTester |
| |Z|,
|Y|, phase angle,
Rp(DCR), Rs(ESR,
DCR), G, X, B,
Cp, Cs, Lp,
Ls, D(dissipation
factor), and Q |
|Z|, |Y|,
phase angle, Rp,
Rs(ESR),
G, X, B, Cp,
Cs, Lp, Ls, D
(dissipation factor), and Q |
|
|
Measurement method |
Source: constant current
10 µA to 100 mA (42 Hz to 1 MHz), 50 µA to 20 mA (1 MHz to 5 MHz), or
constant voltage 10 mV to 5 V (42 Hz to 1 MHz), 50 mV to 1 V (1 MHz to 5
MHz)
Sense: voltage, AC |
Source: constant current
10 µA to 100 mA(AC/DC), or constant voltage 10 mV to 5 V (AC/DC),
Sense: voltage, AC or DC |
| Source
frequency |
42 Hz
to 5 MHz |
DC, or 1 mHz to 100 kHz |
|
Measurement ranges |
|Z|, R,
X: 10.00 m-Ohm to 200.00 m-ohm (depending on
condition), phase angle: -180.00 to +180.00°,
C: 0.3200 pF to 370.00 mF, L: 16.000
nH to 750.00 kH, D: 0.00001 to 9.99999, Q:
0.01 to 999.99,
|Y|, G, B: 5.0000 nS
to 99.999 S |
Z|, R,
X: 10.00 m-Ohm to 200.00 m-ohm (depending on
condition), phase angle: -180.00 to +180.00°,
C: 0.3200 pF to 1.0000 F, L: 16.000 nH
to 750.00 kH, D: 0.00001 to 9.99999, Q:
0.01 to 999.99,
|Y|, G, B: 5.0000 nS
to 99.999 S |
| Basic
accuracy |
|Z|: ±0.08 % rdg. ,
phase angle: ±0.05 ° |
Measurement times
typical values for displaying |Z| |
Fast: 5 msec. to
Slow2: 140 msec. |
Fast: 5 msec. to
Slow2: 828 msec. |
| Display |
99999
full digits, LCD with backlight display |
| Comparator
functions |
Setting: Upper and
lower limit, percentage, or absolute value, Output: 3
levels (Hi, In, Lo), Open-collector, Isolated |
| External
printer |
9442
(use with the
9443-02 or -03/9446/9593-01) |
| Power
supply |
100 to
240 V AC (selectable type), 50/60Hz |
|
Dimensions, mass |
352
mm(13.86 in)W × 124 mm(4.88 in)H × 323 mm(12.72 in)D, 6.5 kg (229.3 oz) |
313
mm(12.32 in)W × 125 mm(4.92 in)H × 290 mm(11.42 in)D, 4.5 kg (158.7 oz) |
|
Accessories |
Power
cord(1), Fuse(1) |
|